56
Step 8
Press the scroll key until the display shows . Send a 10V
signal to terminals 15 and 16 with the correct polarity. Press the
scroll key for at least 3 seconds. The display will blink for a
moment until a new value is obtained. If the display didn’t blink
or if the obtained value is equal to -199.9 or 199.9, then calibra-
tion failed.
• Perform step 9 to calibrate mA function ( if required ) for input
2.
Step 9
Press the scroll key until the display shows . Send a
20mA signal to terminals 15 and 16 with the correct polarity.
Press the scroll key for at least 3 seconds. The display will blink
for a moment until a new value is obtained. If the display didn’t
blink or if the obtained value is equal to -199.9 or 199.9, then
calibration failed.
• Perform step 10 to calibrate offset of cold junction compensa-
tion, if required. Set the DIP switch for T/C input.
Step 10
Set up the equipment according to
the following diagram to calibrate
the cold junction compensation.
Note that a K type thermocouple
must be used.
The 5520A calibrator is configured for K type thermocouple out-
put with internal compensation. Send a 0.00°C signal to the unit
under calibration. The unit under calibration is powered in a still-
air room with a temperature of 25±3°C. Allow at least 20 min-
utes to warm up. The DIP switch is located at the TC input.
Perform step 1 as stated above, then press the scroll key until the
display shows
.
Press the up and down keys until a value
of 0.00 is obtained. Press the scroll key at least 3 seconds. The
display will blink for a moment until a new value is obtained. If
the display didn’t blink or if the obtained value is equal to -5.00
or 40.00, then calibration failed.
• Perform step 11 to calibrate the gain of cold junction compen-
sation if required. If a test chamber for calibration is not avail-
able, perform step 11N to use a nominal value for the cold junc-
tion gain.
Step 11
Set up the equipment the same as in step 10. The unit under cal-
ibration is powered in a still-air room with a temperature of
50±3°C. Allow at least 20 minutes to warm up. The calibrator
source is set at 0.00°C with internal compensation mode.
Perform step 1 as stated above, then press the scroll key until
the display shows
.
Press the up and down keys until a
value of 0.0 is obtained. Press the scroll key for at least 3 sec-
onds. The display will blink for a moment until a new value is
obtained. If the display didn’t blink or if the obtained value is
equal to -199.9 or 199.9, then calibration failed. This setup is
performed in a high temperature chamber, therefore it is recom-
mended to use a computer to perform the procedures.
Step 11N
Perform step 1 as stated above, then press the scroll key until the
display shows . Press the up and down keys until a value
of 0.1 is obtained. Press the scroll key for at least 3 seconds. The
display will blink for a moment until the new value 0.0 is
obtained. Otherwise, calibration failed.
Caution: It is not recommended to use step 11N, since the
cold junction gain is not able to achieve the rated accuracy using
this step.
• Final step
Step 12
Set the DIP switch to your desired position (refer to section 1-3).
Automatic calibration procedures
The programming port (see section 2-18) of the TEC-9300 can
be used for automatic calibration.
The equipment required for automatic calibration is available
upon request.
Manual calibration procedures, continued…